Radial artery pseudoaneurysm complication from use of AO/ASIF volar distal radius plate: A case report
A case of pseudoaneurysm of the radial artery after fixation of a distal radius using the AO/ASIF volar distal radius plate system is presented. This unusual complication resulted from inadequate fracture stabilization and plate preparation. Technical tips on plate shortening and preparation of the...
Gespeichert in:
Veröffentlicht in: | The Journal of hand surgery (American ed.) 2001-05, Vol.26 (3), p.448-453 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A case of pseudoaneurysm of the radial artery after fixation of a distal radius using the AO/ASIF volar distal radius plate system is presented. This unusual complication resulted from inadequate fracture stabilization and plate preparation. Technical tips on plate shortening and preparation of the cut ends are emphasized. (J Hand Surg 2001;26A:448-453. Copyright © 2001 by the American Society for Surgery of the Hand.) |
---|---|
ISSN: | 0363-5023 1531-6564 |
DOI: | 10.1053/jhsu.2001.24138 |