Radial artery pseudoaneurysm complication from use of AO/ASIF volar distal radius plate: A case report

A case of pseudoaneurysm of the radial artery after fixation of a distal radius using the AO/ASIF volar distal radius plate system is presented. This unusual complication resulted from inadequate fracture stabilization and plate preparation. Technical tips on plate shortening and preparation of the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of hand surgery (American ed.) 2001-05, Vol.26 (3), p.448-453
Hauptverfasser: Dao, Khiem D., Venn-Watson, Eric, Shin, Alexander Y.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A case of pseudoaneurysm of the radial artery after fixation of a distal radius using the AO/ASIF volar distal radius plate system is presented. This unusual complication resulted from inadequate fracture stabilization and plate preparation. Technical tips on plate shortening and preparation of the cut ends are emphasized. (J Hand Surg 2001;26A:448-453. Copyright © 2001 by the American Society for Surgery of the Hand.)
ISSN:0363-5023
1531-6564
DOI:10.1053/jhsu.2001.24138