Microwave spectroscopy of myocardial ischemia and infarction. 1. Experimental study

The purpose of this study was to assess whether physiologic changes in canine myocardium due to coronary blood flow reduction, ischemia, and infarction could be detected by changes in dielectrical properties. Measurements were made in the frequency range of 0.2-6.0 GHz. Percent coronary blood flow r...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of biomedical engineering 2000, Vol.28 (1), p.48-54
Hauptverfasser: SEMENOV, S. Y, SVENSON, R. H, TATSIS, G. P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to assess whether physiologic changes in canine myocardium due to coronary blood flow reduction, ischemia, and infarction could be detected by changes in dielectrical properties. Measurements were made in the frequency range of 0.2-6.0 GHz. Percent coronary blood flow reduction was linearly related to the decrease in epsilon" at frequencies of 0.2 GHz(R =-0.997) and 1.1 GHz (R =- 0.9987). In 2 h occlusions, increased conduction time in the infarct area mirrored the temporal changes in dielectrical properties. In 2-week-old infarctions differences, in epsilon' between normal and central infarct zones were statistically significant (P
ISSN:0090-6964
1573-9686
DOI:10.1114/1.253