Measurement of airborne ultrasonic slow waves in calcaneal cancellous bone
Measurements of an airborne ultrasonic wave were made in defatted cancellous bone from the human calcaneus using standard ultrasonic equipment. The wave propagating under these conditions was consistent with a decoupled Biot slow wave travelling in the air alone, as previously reported in gas-satura...
Gespeichert in:
Veröffentlicht in: | Medical engineering & physics 1999-05, Vol.21 (4), p.215-223 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Measurements of an airborne ultrasonic wave were made in defatted cancellous bone from the human calcaneus using standard ultrasonic equipment. The wave propagating under these conditions was consistent with a decoupled Biot slow wave travelling in the air alone, as previously reported in gas-saturated foams. Reproducible measurements of phase velocity and attenuation coefficient were possible, and an estimate of the tortuosity of the trabecular framework was derived from the high frequency limit of the phase velocity. Thus the method offers a new approach to the acoustic characterisation of bone in vitro which, in contrast to existing techniques, has the potential to yield information directly characterising the trabecular structure. |
---|---|
ISSN: | 1350-4533 1873-4030 |
DOI: | 10.1016/S1350-4533(99)00048-X |