Membraneless vaporization unit for direct analysis of solid sample
A new apparatus, called ‘membraneless vaporization’ (MBL-VP) unit was designed and developed for direct analysis of solid samples. Solid analyte was converted into a gaseous form which then reacts with an indicator reagent. Change in absorbance was used to quantitate the analyte. Stirring with a mag...
Gespeichert in:
Veröffentlicht in: | Analytica chimica acta 2007-07, Vol.597 (1), p.157-162 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new apparatus, called ‘membraneless vaporization’ (MBL-VP) unit was designed and developed for direct analysis of solid samples. Solid analyte was converted into a gaseous form which then reacts with an indicator reagent. Change in absorbance was used to quantitate the analyte. Stirring with a magnetic bar was employed to facilitate the evaporation of the gas. Unlike the pervaporation technique, hydrophobic membrane was not required for this MBL-VP technique.
Application of the membraneless technique for direct determination of calcium carbonate in calcium supplements, has shown to be very precise (R.S.D.
=
2.5% for 0.16
mmol
CO
3
2−), with detection limit of 0.5
mg CaCO
3. Results by this method agreed well with flame atomic absorption spectrometric method. Sample throughput was 20
samples
h
−1. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/j.aca.2007.06.036 |