Alignment of the femoral component in a mobile-bearing unicompartmental knee arthroplasty: A study in 10 cadaver femora

Abstract Use of an intramedullary rod is advised for the alignment of the femoral component of an Oxford phase-III prosthesis. There are users moving toward extramedullary alignment, which is merely an indicator of frustration with accuracy of intramedullary alignment. The results of our study with...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The knee 2007-08, Vol.14 (4), p.280-283
Hauptverfasser: Kort, N.P, van Raay, J.J.A.M, Thomassen, B.J.W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Use of an intramedullary rod is advised for the alignment of the femoral component of an Oxford phase-III prosthesis. There are users moving toward extramedullary alignment, which is merely an indicator of frustration with accuracy of intramedullary alignment. The results of our study with 10 cadaver femora demonstrate that use of a short and long intramedullary femoral rod may result in excessive flexion alignment error of the femoral component. Understanding of the extramedullary alignment possibility and experience with the visual alignment of the femoral drill guide is essential toward minimizing potential errors in the alignment of the femoral component.
ISSN:0968-0160
1873-5800
DOI:10.1016/j.knee.2007.04.007