Stereoselective determination of vigabatrin enantiomers in human plasma by high performance liquid chromatography using UV detection
A rapid and simple high-performance liquid chromatographic method for the determination of the R-(−)- and S-(+)-enantiomers of the antiepileptic drug vigabatrin in human plasma is described. After adding the internal standard (1-aminomethyl-cycloheptyl-acetic acid), plasma samples (200 μL) are depro...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2007-07, Vol.854 (1), p.63-67 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A rapid and simple high-performance liquid chromatographic method for the determination of the R-(−)- and S-(+)-enantiomers of the antiepileptic drug vigabatrin in human plasma is described. After adding the internal standard (1-aminomethyl-cycloheptyl-acetic acid), plasma samples (200
μL) are deproteinized with acetonitrile and the supernatant is derivatized with 2,4,6 trinitrobenzene sulfonic acid (TNBSA). Separation is achieved on a reversed-phase cellulose-based chiral column (Chiralcel-ODR, 250
mm
×
4.6
mm i.d.) using 0.05
M potassium hexafluorophosphate (pH 4.5)/acetonitrile/ethanol (50:40:10
vol/vol/vol) as mobile phase at a flow-rate of 0.9
mL/min. Chromatographic selectivity is improved by concentrating the derivatives on High Performance Extraction Disk Cartridges prior to injection. Detection is at 340
nm. Calibration curves are linear (
r
2
≥
0.999) over the range of 0.5–40
μg/mL for each enantiomer, with a limit of quantification of 0.5
μg/mL for both analytes. The assay is suitable for therapeutic drug monitoring and for single-dose pharmacokinetic studies in man. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2007.03.042 |