Simultaneous assay of metformin and glibenclamide in human plasma based on extraction-less sample preparation procedure and LC/(APCI)MS
Separation of metformin and glibenclamide was achieved within a single chromatographic run on a Zorbax CN column, under isocratic conditions, using acetonitrile and aqueous component (0.01 moles/L ammonium acetate adjusted at pH 3.5 with acetic acid) in volumetric ratio 1/1. Plasma sample preparatio...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2007-07, Vol.854 (1), p.211-218 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Separation of metformin and glibenclamide was achieved within a single chromatographic run on a Zorbax CN column, under isocratic conditions, using acetonitrile and aqueous component (0.01
moles/L ammonium acetate adjusted at pH 3.5 with acetic acid) in volumetric ratio 1/1. Plasma sample preparation is based on protein precipitation by means of organic solvent addition. 1,3,5-Triazine-2,4,6-triamine (IS1) was used as internal standard for metformin, while gliquidone (IS2) played the same role for glibenclamide. Detection was performed with an ion trap mass analyzer, using atmospheric pressure chemical ionization (APCI). A single MS stage was used for detection of metformin and IS1, by extracting ion chromatograms corresponding to molecular ions. MS/MS detection in the SRM mode was used for glibenclamide (
m/
z transition from 494 to 369
Da) and IS2 (
m/
z transition from 528 to 403
Da). The method produces linear responses up to 2000
ng/mL for metformin and 400
ng/mL for glibenclamide, respectively. Low limits of quantification were found in the 40
ng/mL range for metformin and at the 4
ng/mL level for glibenclamide. Precision was characterized by relative standard deviations (RSD%) below 9%. The analytical method was successfully applied to a single dose, open-label, randomized, two-period, two-sequence, crossover bioequivalence study of two commercially available anti-diabetic combinations containing 400
mg metformin and 2.5
mg of glibenclamide per coated tablet. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2007.04.032 |