Miniaturized solid-phase extraction with resin disks

The properties and advantages of membrane disks for solid-phase extraction (SPE) are described. Miniaturization is a trend in SPE, as well as chemical analysis in general. A semimicro method is reviewed in which an extraction disk 4 mm in diameter is used for SPE. Even smaller scale separations are...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Chromatography A 2001-02, Vol.909 (1), p.79-85
Hauptverfasser: Fritz, James S., Masso, Jeremy J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The properties and advantages of membrane disks for solid-phase extraction (SPE) are described. Miniaturization is a trend in SPE, as well as chemical analysis in general. A semimicro method is reviewed in which an extraction disk 4 mm in diameter is used for SPE. Even smaller scale separations are possible with a device in which a membrane 0.7 mm in diameter was incorporated into the needle of a 50-μl syringe. Aqueous samples containing 10 ppb of eight substituted benzenes were passed through the miniaturized-SPE syringe automatically using a single-syringe infusion pump. Elution volumes of 5 μl provided 500-fold concentrations and delivered average recoveries greater than 90% and an average relative standard deviation (RSD) of 4.6% for the analytes. Direct injection of the 5 μl eluate from the miniaturized-SPE syringe into a gas chromatograph also produced average recoveries greater than 90% and an average RSD of 6.2%.
ISSN:0021-9673
DOI:10.1016/S0021-9673(00)01014-1