Prevention of surface encrustation of urological implants by coating with inhibitors

The encrustation of materials used for urological implants is as yet an unresolved problem. The crystallisation-inhibiting effect of the glycosaminoglycan heparin was used to reduce encrustation. Heparin was covalently bound to the surface of slotted-tube stents of tantalum and stainless steel using...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomaterials 2001-03, Vol.22 (5), p.503-507
Hauptverfasser: Hildebrandt, P, Sayyad, M, Rzany, A, Schaldach, M, Seiter, H
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The encrustation of materials used for urological implants is as yet an unresolved problem. The crystallisation-inhibiting effect of the glycosaminoglycan heparin was used to reduce encrustation. Heparin was covalently bound to the surface of slotted-tube stents of tantalum and stainless steel using a spacer molecule. To verify the inhibition of crystallisation processes, reproducible in vitro tests and in vivo tests using the rat as animal model were carried out. The in vitro and in vivo experiments showed that the heparin coating has a significant influence on the encrustation of the surface. After 7 days in vitro and 120 days in vivo, heparin coated stents were free of encrustation, whereas the uncoated reference stents were extensively covered.
ISSN:0142-9612
1878-5905
DOI:10.1016/S0142-9612(00)00217-9