Measurement of Trace Elements in Proteins Extracted from Liver by Size Exclusion Chromatography-Inductively Coupled Plasma-Mass Spectrometry with a Magnetic Sector Mass Spectrometer
Proteins are extracted from liver into aqueous buffer at pH 7 and separated by size exclusion chromatography (SEC). Inductively coupled plasma mass spectrometry (ICP-MS) with a magnetic sector mass spectrometer is used to identify those protein fractions that contain Cu, Zn, Mn, Fe, Cd, S, P, Mo, Co...
Gespeichert in:
Veröffentlicht in: | Analytical biochemistry 2001-01, Vol.288 (1), p.89-96 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Proteins are extracted from liver into aqueous buffer at pH 7 and separated by size exclusion chromatography (SEC). Inductively coupled plasma mass spectrometry (ICP-MS) with a magnetic sector mass spectrometer is used to identify those protein fractions that contain Cu, Zn, Mn, Fe, Cd, S, P, Mo, Co, Ca, or Mg. The experimental setup provides very high sensitivity. Measurements at medium spectral resolution remove polyatomic interferences for some difficult elements like Fe, S, and P. Some elements are found in different molecular weight proteins; for example, cadmium binds to four different protein fractions (>400 kDa, 70 kDa, and metallothionein). Other elements like Mo, Ca, and Mg are present only in low-molecular-weight proteins or other small molecules. |
---|---|
ISSN: | 0003-2697 1096-0309 |
DOI: | 10.1006/abio.2000.4885 |