Online Preconcentration by Transient Isotachophoresis in Linear Polymer on a Poly(methyl methacrylate) Microchip for Separation of Human Serum Albumin Immunoassay Mixtures
Online preconcentration of human serum albumin (HSA) and its immunocomplex with a monoclonal antibody by on-chip transient isotachophoresis is reported. An 800-fold signal enhancement was achieved following the preconcentration on standard cross-channel microchips made of poly (methyl methacrylate)....
Gespeichert in:
Veröffentlicht in: | Analytical chemistry (Washington) 2007-05, Vol.79 (10), p.3667-3672 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Online preconcentration of human serum albumin (HSA) and its immunocomplex with a monoclonal antibody by on-chip transient isotachophoresis is reported. An 800-fold signal enhancement was achieved following the preconcentration on standard cross-channel microchips made of poly (methyl methacrylate). Sample injection, preconcentration, and separation were done continuously and controlled solely by a sequential voltage switching program. The preconcentration was followed by on-chip nondenaturing gel electrophoresis in methylcellulose solution. The method was applied to microchip electrophoresis immunoassay of HSA. Baseline separation of HSA and its immunocomplex was achieved in 25 s in the first 1 cm of the microchannel. In a direct immunoassay, the minimum detectable concentration of fluorescent labeled HSA by laser-induced fluorescence detection was 7.5 pM. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/ac0623890 |