In vitro evaluation of the CeVOX continuous central venous oxygenation monitoring system

We investigated the in vitro performance of the CeVOX system for continuous monitoring of central venous oxygen saturation by spectrophotometry (Pulsion Medical Systems, Munich, Germany). Oxygen inflow into the system was varied, and oxygen saturation values measured by CeVOX were documented. Blood...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 2008-04, Vol.63 (4), p.412-417
Hauptverfasser: Baulig, W, Dullenkopf, A, Hasenclever, P, Schmid, E R, Weiss, M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We investigated the in vitro performance of the CeVOX system for continuous monitoring of central venous oxygen saturation by spectrophotometry (Pulsion Medical Systems, Munich, Germany). Oxygen inflow into the system was varied, and oxygen saturation values measured by CeVOX were documented. Blood samples were simultaneously taken to assess oxygen saturation by co-oximetry, and values were compared by Bland-Altman analysis. Sixty-six data pairs were obtained at CeVOX and co-oximetry values of 16-99% and 5.5-100%, respectively. Overall, CeVOX values only slightly overestimated co-oximetry values (mean bias +2.4%), but limits of agreement (2 SD of bias) were wide (-11.8 to +16.6%). Saturation measured by CeVOX underestimated that measured by co-oximetry at higher oxygen concentrations and overestimated it at lower oxygen concentrations. There was a nearly linear correlation of the mean bias, suggesting a systematic error. We conclude that the current version of the CeVOX system does not reliably reflect oxygen saturation.
ISSN:1365-2044
DOI:10.1111/j.1365-2044.2007.05376.x