Investigation of high-speed gas chromatography using synchronized dual-valve injection and resistively heated temperature programming
High-speed temperature programming is implemented via the direct resistive heating of the separation column (2.3 m MXT-5 Silicosteel column with a 180 μm I.D. and a 0.4 μm 5% phenyl/95% dimethyl polysiloxane film). Resistive temperature programming was coupled with synchronized dual-valve injection...
Gespeichert in:
Veröffentlicht in: | Journal of Chromatography A 2007-05, Vol.1148 (2), p.236-243 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | High-speed temperature programming is implemented via the direct resistive heating of the separation column (2.3
m MXT-5 Silicosteel column with a 180
μm I.D. and a 0.4
μm 5% phenyl/95% dimethyl polysiloxane film). Resistive temperature programming was coupled with synchronized dual-valve injection (with an injection pulse width of 2
ms), producing a complete high-speed gas chromatography (GC) system. A comparison of isothermal and temperature programmed separations of seven
n-alkanes (C
6 and C
8–C
13) shows a substantial improvement of peak width and peak capacity with temperature programming. The system was further implemented in separations of a mixture of analytes from various chemical classes.
Separations of the
n-alkane mixture using three different temperature programming rates are reported. A temperature programming rate as high as 240
°C/s is demonstrated. The method for determination of temperature programming rate, based on isothermal data, is discussed. The high-speed resistive column heating temperature programming resulted in highly reproducible separations. The highest rate of temperature programming (240
°C/s) resulted in retention time and peak width RSD, on average, of 0.5 and 1.4%, respectively, for the
n-alkane mixture. This high level of precision was achieved with peak widths-at-half-height ranging from 13 to 36
ms, and retention times ranging from 147 to 444
ms (for
n-hexane to
n-tridecane). |
---|---|
ISSN: | 0021-9673 |
DOI: | 10.1016/j.chroma.2007.03.029 |