Solid state biliary manometry catheter: impact on diagnosis and post-study pancreatitis

Perfusion manometry of the sphincter of Oddi has been the standard for the investigation of patients with presumed sphincter of Oddi dysfunction (SOD). Microtransducer manometry (MTM) of the sphincter of Oddi represents an alternative to perfusion manometry. The technical success and reproducibility...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Current gastroenterology reports 2007-04, Vol.9 (2), p.171-174
Hauptverfasser: Frenz, Markus B, Wehrmann, Till
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Perfusion manometry of the sphincter of Oddi has been the standard for the investigation of patients with presumed sphincter of Oddi dysfunction (SOD). Microtransducer manometry (MTM) of the sphincter of Oddi represents an alternative to perfusion manometry. The technical success and reproducibility of MTM are as good as for perfusion manometry. Current data suggest that the upper limit for normal of basal sphincter of Oddi pressures measured with MTM lies at approximately 35 mm Hg. Pancreatitis risk after MTM in patients with SOD compares favorably with that after perfusion manometry. Low cost and ease of handling make MTM of the sphincter of Oddi an attractive alternative.
ISSN:1522-8037
1534-312X
DOI:10.1007/s11894-007-0013-4