Evaluation of cardiac assays on a benchtop chemiluminescent enzyme immunoassay analyzer, PATHFAST

Devices for cardiac immunoassay are widely available at point of care settings. The lateral flow method is the most popular solution for ease of use. But its major shortcomings, poor assay precision, and low analytical sensitivity may lead to false negative results. Therefore, confirmatory testing b...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical biochemistry 2008-04, Vol.375 (1), p.144-146
Hauptverfasser: Kurihara, Takashi, Yanagida, Atsushi, Yokoi, Hiroyuki, Koyata, Atsushi, Matsuya, Takeshi, Ogawa, Junichi, Okamura, Yoshikazu, Miyamoto, Daishi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Devices for cardiac immunoassay are widely available at point of care settings. The lateral flow method is the most popular solution for ease of use. But its major shortcomings, poor assay precision, and low analytical sensitivity may lead to false negative results. Therefore, confirmatory testing by a routine lab analyzer sometimes is necessary. In the current study, we evaluated the cardiac assays troponin I, MB isoenzyme of creatine kinase, myoglobin, and N-terminal pro brain natriuretic peptide on a chemiluminescent enzyme immunoassay analyzer, PATHFAST. All of the assays demonstrated correlation coefficients higher than 0.97 against the predicate devices along with good total precision (coefficients of variation
ISSN:0003-2697
1096-0309
DOI:10.1016/j.ab.2007.12.030