Development and validation of a reversed-phase ion-pair high-performance liquid chromatographic method for the determination of risedronate in pharmaceutical preparations
A stability indicating, reversed-phase ion-pair high-performance liquid chromatographic method was developed and validated for the determination of risedronate in pharmaceutical dosage forms. The determination was performed on a BDS C 18 analytical column (250 mm × 4.6 mm i.d., 5 μm particle size);...
Gespeichert in:
Veröffentlicht in: | Analytica chimica acta 2007-02, Vol.584 (1), p.153-159 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A stability indicating, reversed-phase ion-pair high-performance liquid chromatographic method was developed and validated for the determination of risedronate in pharmaceutical dosage forms. The determination was performed on a BDS C
18 analytical column (250
mm
×
4.6
mm i.d., 5
μm particle size); the mobile phase consisted of 0.005
M tetrabutylammonium hydroxide and 0.005
M pyrophosphate sodium (pH 7.0) mixed with acetonitrile in a ratio (78:22, v/v) and pumped at a flow rate 1.00
mL
min
−1. The ultraviolet (UV) detector was operated at 262
nm. The retention times of magnesium ascorbyl phosphate, which was used as internal standard and risedronate were 4.94 and 5.95
min, respectively. The calibration graph was ranged from 2.50 to 20.00
μg
mL
−1, while detection and quantitation limits were found to be 0.48 and 1.61
μg
mL
−1, respectively. The intra- and inter-day percentage relative standard deviations, %R.S.D., were less than 5.9%, while the relative percentage error, %
E
r, was less than 0.4%. The method was applied to the quality control of commercial tablets and content uniformity test and proved to be suitable for rapid and reliable quality control. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/j.aca.2006.11.012 |