Using coupled monolithic rods for ultra-high peak capacity LC and LC-MS under normal LC operating pressures
Up to ten individual reversed-phase monolithic LC columns were efficiently coupled to equal a total column length of 1 m. The column generated up to 90,000 theoretical plates under optimum conditions with a peak capacity under gradient conditions of 244 peaks h(-1) at a total column backpressure of...
Gespeichert in:
Veröffentlicht in: | Analyst (London) 2008-02, Vol.133 (2), p.180-183 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Up to ten individual reversed-phase monolithic LC columns were efficiently coupled to equal a total column length of 1 m. The column generated up to 90,000 theoretical plates under optimum conditions with a peak capacity under gradient conditions of 244 peaks h(-1) at a total column backpressure of just 140 bar, directly coupled to ESI-MS. |
---|---|
ISSN: | 0003-2654 1364-5528 |
DOI: | 10.1039/b714807k |