Mass spectrometric and physiological validation of a sensitive, automated, direct immunoassay for serum estradiol using the Architect

Measurement of estradiol (E 2) plays a critical role in the diagnosis and clinical management of reproductive disorders. The challenge for all currently available direct methods for measuring E 2 is to provide accuracy and precision across a wide dynamic range. We describe the development and multi-...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinica chimica acta 2008-02, Vol.388 (1), p.99-105
Hauptverfasser: Sluss, Patrick M., Hayes, Frances J., Adams, Judith M., Barnes, Wayne, Williams, Gregg, Frost, Stephen, Ramp, John, Pacenti, David, Lehotay, Denis C., George, Sabu, Ramsay, Carol, Doss, Robert C., Crowley, William F.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Measurement of estradiol (E 2) plays a critical role in the diagnosis and clinical management of reproductive disorders. The challenge for all currently available direct methods for measuring E 2 is to provide accuracy and precision across a wide dynamic range. We describe the development and multi-site performance evaluation of a direct E 2 assay on the Architect i2000®. Assay performance and method comparisons were performed by testing specimens from men, healthy women with regular menstrual cycles, and post-menopausal women using the Architect assay and isotope dilution, gas chromatography-mass spectrometry (ID/GC-MS). Reference intervals were established by testing prospectively collected daily blood draws from 42 healthy women, 72 postmenopausal women and 101 males. No unexpected cross-reactivity or interference was observed for over 40 compounds tested. Recovery was 100 ± 10% in the presence of estrone and estriol. Functional sensitivity (%CV < 20%) was < 15 pg/ml. 1 1 To convert conventional pg/ml [ng/l] to SI units [pmol/l] multiply by 3.671. The imprecision of the assay was < 7.1% (total CV), < 2.5%, and < 2.3% for control sera containing 45, 190, and 600 pg/ml estradiol, respectively. The assay had a correlation of y = 1.033 x + 0.3156, r 2 = 0.99, n = 131 compared to ID/GC-MS. Reference intervals for the current Architect Estradiol assay are reported. Format changes resulted in dramatic improvement in the performance and accuracy of this direct, fully automated assay. The assay is standardized by ID/GC-MS. The assay is clinically useful for serum concentrations from 15 to > 4000 pg/ml.
ISSN:0009-8981
1873-3492
DOI:10.1016/j.cca.2007.10.020