Automatic blood pressure measurement: the oscillometric waveform shape is a potential contributor to differences between oscillometric and auscultatory pressure measurements

OBJECTIVETo explore the differences between oscillometric and auscultatory measurements. METHODFrom a simulator evaluation of a non-invasive blood pressure (NIBP) device regenerating 242 oscillometric blood pressure waveforms from 124 subjects, 10 waveforms were selected based on the differences bet...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of hypertension 2008-01, Vol.26 (1), p.35-43
Hauptverfasser: Amoore, John N, Lemesre, Yann, Murray, Ian C, Mieke, Stephan, King, Susan T, Smith, Fiona E, Murray, Alan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:OBJECTIVETo explore the differences between oscillometric and auscultatory measurements. METHODFrom a simulator evaluation of a non-invasive blood pressure (NIBP) device regenerating 242 oscillometric blood pressure waveforms from 124 subjects, 10 waveforms were selected based on the differences between the NIBP (oscillometric) and auscultatory pressure measurements. Two waveforms were selected for each of five criteriasystolic over and underestimation; diastolic over and underestimation; and close agreement for both systolic and diastolic pressures. The 10 waveforms were presented to seven different devices and the oscillometric–auscultatory pressure differences were compared between devices and with the oscillometric waveform shapes. RESULTSConsistent patterns of waveform-dependent over and underestimation of systolic and diastolic pressures were shown for all seven devices. The mean and standard deviation, for all devices, of oscillometric–auscultatory pressure differences werefor the systolic overestimated waveforms, 36 ± 28/−6 ± 3 and 23 ± 2/−1 ± 3 mmHg (systolic/diastolic differences); for systolic underestimated waveforms, −21 ± 5/−4 ± 3 and −11 ± 4/−3 ± 3 mmHg; for diastolic overestimated waveforms, 3 ± 4/12 ± 5 and 17 ± 6/10 ± 2 mmHg; for diastolic underestimated waveforms, 1 ± 4/−22 ± 4 and −9 ± 6/−29 ± 4 mmHg; and for the two waveforms with good agreement, 0 ± 6/0 ± 3 and −2 ± 4/−4 ± 3 mmHg. Waveforms for which devices showed good oscillometric and auscultatory agreement had smooth envelopes with clearly defined peaks, compared with the broader plateau and complex shapes of those waveforms for which devices over or underestimated pressures. CONCLUSIONBy increasing the understanding of the characteristics and limitations of the oscillometric method and the effects of waveform shape on pressure measurements, simulator evaluation should lead to improvements in NIBP devices.
ISSN:0263-6352
1473-5598
DOI:10.1097/HJH.0b013e3282f06ec7