An immunological Interleukine-6 capacitive biosensor using perturbation with a potentiostatic step
An instrument for potentiostatic capacitance measurements, based on perturbation with a potentiostatic step was used. The capacitive sensor consisted of self-assembled sulfur compounds on gold to which antibodies towards Interleukine 6, Il-6, had been immobilized. The biosensor was part of a potenti...
Gespeichert in:
Veröffentlicht in: | Biosensors & bioelectronics 1998-11, Vol.13 (10), p.1061-1068 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An instrument for potentiostatic capacitance measurements, based on perturbation with a potentiostatic step was used. The capacitive sensor consisted of self-assembled sulfur compounds on gold to which antibodies towards Interleukine 6, Il-6, had been immobilized. The biosensor was part of a potentiostatic three-electrode system with an extra reference electrode. Two different methods using epoxy- or carbodiimide coupling of the polyclonal antibodies were compared. The antigen Il-6 could be detected from 5×10
−16 to 5×10
−13
M with one immobilization method and to more than 5×10
−9
M with the other. No labels were necessary since the binding of the antigen was detected directly. The insulating properties of the different layers of the biosensor were investigated. |
---|---|
ISSN: | 0956-5663 1873-4235 |