Analytical Validation and Biological Evaluation of a High Molecular-Weight Adiponectin ELISA
Of the 3 circulating multimeric forms of adiponectin, the high-molecular-weight (HMW) form, as measured by size-exclusion and/or immunoblotting techniques, is a better index of insulin sensitivity for monitoring health and disease than is total adiponectin. We aimed to develop a simple ELISA to meas...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 2007-12, Vol.53 (12), p.2144-2151 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Of the 3 circulating multimeric forms of adiponectin, the high-molecular-weight (HMW) form, as measured by size-exclusion and/or immunoblotting techniques, is a better index of insulin sensitivity for monitoring health and disease than is total adiponectin. We aimed to develop a simple ELISA to measure HMW adiponectin.
We pretreated serum or plasma samples with digestion solution containing proteinase K (Millipore, ESDS). HMW (Millipore, EZHMWA-64K) and total adiponectin (Millipore, EZHADP-61K) concentrations were measured in treated and untreated samples, respectively, from 108 individuals and from 20 morbidly obese patients before and at 1, 3, 6, and 12 months after gastric-bypass surgery.
The ELISA has a dynamic range of 3-200 microg/L and a detection limit of 0.8 microg/L. Intraassay and interassay CVs were |
---|---|
ISSN: | 0009-9147 1530-8561 |
DOI: | 10.1373/clinchem.2007.090670 |