A Newly Designed Resonance Frequency Analysis Device for Dental Implant Stability Detection

Resonance frequency (RF) analysis technology was used to design a new dental implant stability detector. To calibrate and test the performance of this novel apparatus, in vitro and in vivo models, respectively, were used. The RF values of the test implants detected using our new device and a commerc...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Dental Materials Journal 2007, Vol.26(5), pp.665-671
Hauptverfasser: CHANG, Wei-Jen, LEE, Sheng-Yang, WU, Chen-Che, LIN, Che-Tong, ABIKO, Yoshimitsu, YAMAMICHI, Nobuyuki, HUANG, Haw-Ming
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Resonance frequency (RF) analysis technology was used to design a new dental implant stability detector. To calibrate and test the performance of this novel apparatus, in vitro and in vivo models, respectively, were used. The RF values of the test implants detected using our new device and a commercially available analogous device (Osstell) were compared. Further, implant stability status was also detected clinically using our device at 2, 4, 8, and 12 weeks after surgery. A high correlation was demonstrated between the values measured with the two devices (y=0.31x-12.45; R2=0.98, p
ISSN:0287-4547
1881-1361
DOI:10.4012/dmj.26.665