Comparison of immediate and follow-up results of the short and long NIR stent with the palmaz-schatz stent
The intrinsic characteristics of a stent including stent length may affect both procedural success and long-term outcome. The present study evaluated the immediate and follow-up results after implantation of the short and long NIR stent and compared these results with the Palmaz-Schatz stent. Betwee...
Gespeichert in:
Veröffentlicht in: | The American journal of cardiology 1999-09, Vol.84 (5), p.499-504 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The intrinsic characteristics of a stent including stent length may affect both procedural success and long-term outcome. The present study evaluated the immediate and follow-up results after implantation of the short and long NIR stent and compared these results with the Palmaz-Schatz stent. Between July 1995 and December 1996, stenting with a 16-mm NIR stent (NIR-16), a 32-mm NIR stent (NIR-32), or a Palmaz-Schatz stent (PS) was performed in 68, 57, and 155 lesions, respectively. There were no significant differences in the incidences of delivery failure (PS, 2.6%: NIR-16, 4.4%; NIR-32, 5.3%; p = NS) and procedural success (PS, 92%; NIR-16, 93%; NIR-32, 93%; p = NS) among the 3 groups. The reference vessel diameter was smaller in lesions with a 32-mm NIR stent than in those with a Palmaz-Schatz stent (PS, 3.14 ± 0.58, NIR-16, 3.00 ± 0.50; NIR-32, 2.90 ± 0.47 mm; p |
---|---|
ISSN: | 0002-9149 1879-1913 |
DOI: | 10.1016/S0002-9149(99)00366-5 |