The effect of steam sterilisation at 134 degrees C on light intensity provided by fibrelight Macintosh laryngoscopes

We studied the repeated effect of sterilisation on light intensity in laryngoscopes from Penlon, Riester, Heine (two different blades), Medicon and Upsher. Light intensity was measured by a light meter using two methods. Measurements were performed before the decontamination procedure was carried ou...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 1999-09, Vol.54 (9), p.875-878
Hauptverfasser: Bucx, M J, Veldman, D J, Beenhakker, M M, Koster, R
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We studied the repeated effect of sterilisation on light intensity in laryngoscopes from Penlon, Riester, Heine (two different blades), Medicon and Upsher. Light intensity was measured by a light meter using two methods. Measurements were performed before the decontamination procedure was carried out and subsequently after each series of 25 procedures until a total of 200 cycles was reached. Using method 1 (and 2), the reduction in light intensity after 200 cycles was 100% (100%; no light emitted), 37% (13%), 75% (69%), 79% (60%), 37% (14%) and 63% (55%) for each blade, respectively.
ISSN:0003-2409