The aim of this study was to clarify the acute performance of steroid-eluting screw-in leads in comparison with that of nonsteroid screw-in leads for atrial free wall pacing. In 114 cases (68 males, 46 females, average age 70 years) with atrial free wall pacing by screw-in leads, pacing thresholds a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:JAPANESE CIRCULATION JOURNAL 1999, Vol.63(7), pp.514-516
Hauptverfasser: Nakazato, Yuji, Nakata, Yasuro, Mineta, Yoriaki, Nakazato, Kaoru, Yasuda, Masayuki, Sumiyoshi, Masataka, Yamaguchi, Hiroshi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The aim of this study was to clarify the acute performance of steroid-eluting screw-in leads in comparison with that of nonsteroid screw-in leads for atrial free wall pacing. In 114 cases (68 males, 46 females, average age 70 years) with atrial free wall pacing by screw-in leads, pacing thresholds and P-wave amplitudes were compared at the time of implantation and 1 week later between 68 cases of nonsteroid and 46 cases of steroid-eluting screw-in leads. No significant differences were seen between the 2 groups at implantation in either voltage or current thresholds measured at pulse widths of 0.1, 0.3, 0.6, 1.0, 2.0 ms, or P-wave amplitudes. Pulse width thresholds at outputs of 2.5 V and 5.0 V were significantly lower for steroid leads 1 week after implantation (2.5 V: 0.34±0.27 ms nonsteroid vs 0.12±0.08 ms steroid, p