Sensitive method for the determination of baclofen in plasma by means of solid-phase extraction and liquid chromatography–tandem mass spectrometry
A simple and sensitive method for the determination of baclofen in plasma is described. Baclofen and the internal standard, KM 08205, were isolated from plasma by solid-phase extraction using C 18 material. After separation by reversed-phase liquid chromatography, the analytes were detected with tan...
Gespeichert in:
Veröffentlicht in: | Journal of Chromatography A 1999-06, Vol.846 (1), p.169-173 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple and sensitive method for the determination of baclofen in plasma is described. Baclofen and the internal standard, KM 08205, were isolated from plasma by solid-phase extraction using C
18 material. After separation by reversed-phase liquid chromatography, the analytes were detected with tandem mass spectrometry. The extraction procedure was optimised regarding the solid-phase extraction material, the pH in the conditioning solution and the washing volume. The method was proven to be selective and sensitive with an absolute recovery of about 95%, a relative standard deviation below 5% and a limit of quantification of 10 nmol/l. |
---|---|
ISSN: | 0021-9673 |
DOI: | 10.1016/S0021-9673(99)00013-8 |