Sensitive method for the determination of baclofen in plasma by means of solid-phase extraction and liquid chromatography–tandem mass spectrometry

A simple and sensitive method for the determination of baclofen in plasma is described. Baclofen and the internal standard, KM 08205, were isolated from plasma by solid-phase extraction using C 18 material. After separation by reversed-phase liquid chromatography, the analytes were detected with tan...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Chromatography A 1999-06, Vol.846 (1), p.169-173
Hauptverfasser: Flärdh, Maria, Jacobson, Britt-Marie
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A simple and sensitive method for the determination of baclofen in plasma is described. Baclofen and the internal standard, KM 08205, were isolated from plasma by solid-phase extraction using C 18 material. After separation by reversed-phase liquid chromatography, the analytes were detected with tandem mass spectrometry. The extraction procedure was optimised regarding the solid-phase extraction material, the pH in the conditioning solution and the washing volume. The method was proven to be selective and sensitive with an absolute recovery of about 95%, a relative standard deviation below 5% and a limit of quantification of 10 nmol/l.
ISSN:0021-9673
DOI:10.1016/S0021-9673(99)00013-8