Comparison of high-pressure liquid chromatography (HPLC) and Griess reagent-spectroscopic methods for the measurement of nitrate in serum from healthy individuals in the Nordic countries

Bioavailability of NO can be estimated by measuring the concentration of nitrate (NO 3) in serum. However, the methods used for the measurement NO 3 in plasma or serum show a great degree of variation. Therefore, we compared two analytical methods for the measurement of NO 3 in serum. The concentrat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical biochemistry 2008-12, Vol.41 (18), p.1474-1481
Hauptverfasser: Larsen, Tine Lise, Nilsen, Valentina, Andersen, Dag Olav, Francis, George, Rustad, Pål, Mansoor, Mohammad Azam
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Bioavailability of NO can be estimated by measuring the concentration of nitrate (NO 3) in serum. However, the methods used for the measurement NO 3 in plasma or serum show a great degree of variation. Therefore, we compared two analytical methods for the measurement of NO 3 in serum. The concentration of NO 3 in 600 serum samples collected from healthy individuals was determined by the HPLC and by the Griess reagent-spectroscopic method. The concentration of NO 3 in the samples was 29.4 ± 16.1 μmol/L and 26.2 ± 14.0 μmol/L (mean ± SD) measured by HPLC and Griess reagent-spectroscopic method respectively ( p < 0.0001). We detected a significant correlation between the two methods ( R = 0.81, p < 0.0001). A significant correlation between the two methods may suggest that either method can be used for the measurement of NO 3 in serum, however the Griess reagent-spectroscopic method measures lower concentrations of NO 3 than the HPLC method.
ISSN:0009-9120
1873-2933
DOI:10.1016/j.clinbiochem.2008.08.083