Measurement of Plasma Renin Activity with Use of HPLC-Electrospray-Tandem Mass Spectrometry
The measurement of renin activity is complicated by difficulties in the quantification of angiotensin 1 (Ang1), the product of the renin-catalyzed reaction. We report an HPLC-electrospray-tandem mass spectrometry (HPLC-ESI-MS/MS) method for the quantification of Ang1 as a measure of plasma renin act...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 1999-05, Vol.45 (5), p.659-664 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The measurement of renin activity is complicated by difficulties in the quantification of angiotensin 1 (Ang1), the product of the renin-catalyzed reaction. We report an HPLC-electrospray-tandem mass spectrometry (HPLC-ESI-MS/MS) method for the quantification of Ang1 as a measure of plasma renin activity (PRA).
After incubation (37 degrees C for 3 or 18 h), samples were prepared using C18 solid-phase extraction. [Val]5Ang1 was used as the internal standard (IS). Chromatography was performed on a C18 column, using 200 mL/L ammonium acetate buffer-800 mL/L methanol as the mobile phase. The flow rate was 150 microL/min, with a chromatographic run time of 5 min/sample. Mass spectrometric detection was in the positive ionization mode with selected reaction monitoring (Ang1 m/z 649.0-->784.0; IS m/z 641.9-->770.4).
The assay was linear over the range 2.5-500 ng Ang1/mL, which corresponded to a limit of detection (signal-to-noise ratio of 3:1) of PRA of 0.14 ng Ang1. mL-1. h-1. The imprecision (CV) of the assay at PRA values of 26.1, 13.5, 3.2, and 0.78 ng Ang1. mL-1. h-1 was 7.0%, 7.0%, 15%, and 11%, respectively. Absolute recoveries were 92.3% (Ang1) and 87.4% (IS). Incubation times of 3 h vs 18 h in the PRA assay gave good agreement at PRA |
---|---|