Quality Assurance in Mammography: Artifact Analysis
Evaluation of mammograms for artifacts is essential for mammographic quality assurance. A variety of mammographic artifacts (ie, variations in mammographic density not caused by true attenuation differences) can occur and can create pseudolesions or mask true abnormalities. Many artifacts are readil...
Gespeichert in:
Veröffentlicht in: | Radiographics 1999-03, Vol.19 (2), p.503-522 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Evaluation of mammograms for artifacts is essential for mammographic quality assurance. A variety of mammographic artifacts
(ie, variations in mammographic density not caused by true attenuation differences) can occur and can create pseudolesions
or mask true abnormalities. Many artifacts are readily identified, whereas others present a true diagnostic challenge. Factors
that create artifacts may be related to the processor (eg, static, dirt or excessive developer buildup on the rollers, excessive
roller pressure, damp film, scrapes and scratches, incomplete fixing, power failure, contaminated developer), the technologist
(eg, improper film handling and loading, improper use of the mammography unit and related equipment, positioning and darkroom
errors), the mammography unit (eg, failure of the collimation mirror to rotate, grid inhomogeneity, failure of the reciprocating
grid to move, material in the tube housing, compression failure, improper alignment of the compression paddle with the Bucky
tray, defective compression paddle), or the patient (eg, motion, superimposed objects or substances [jewelry, body parts,
clothing, hair, implanted medical devices, foreign bodies, substances on the skin]). Familiarity with the broad range of artifacts
and the measures required to eliminate them is vital. Careful attention to darkroom cleanliness, care in film handling, regularly
scheduled processor maintenance and chemical replenishment, daily quality assurance activities, and careful attention to detail
during patient positioning and mammography can reduce or eliminate most mammographic artifacts. |
---|---|