Estimation of Radial Strain and Rotation Using a New Algorithm Based on Speckle Tracking
Background The aim of this study was to test the ability of a new algorithm to accurately measure point-to-point Lagrangian strain (LS) and local rotation (ROT). Change in distance between 2 separate regions of interest (ROIs) can theoretically be computed with speckle tracking (SpT) and used to cal...
Gespeichert in:
Veröffentlicht in: | Journal of the American Society of Echocardiography 2008-10, Vol.21 (10), p.1168-1174 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background The aim of this study was to test the ability of a new algorithm to accurately measure point-to-point Lagrangian strain (LS) and local rotation (ROT). Change in distance between 2 separate regions of interest (ROIs) can theoretically be computed with speckle tracking (SpT) and used to calculate LS in any tissue location with angle independence and high spatial resolution. Similarly, tracking an ROI relative to a fixed point should provide an estimate of ROT. Methods Two dynamic phantoms (60 beats/min) were scanned in short axis at frame rates of 30, 60, and 90 Hz. To estimate LS, 2 ROIs were positioned immediately beneath the inner and outer borders of the superior wall of the first phantom and tracked using SpT. LS derived from SpT (SpT-LS) was compared with LS measured by sonomicrometers placed on the inner and outer walls of the phantom (SN-LS). To estimate ROT, the rotational vectors around the centroid of a second phantom were calculated for 3 epicardial bead targets imaged with gated computed tomography (CT) and compared with measurements derived from SpT. Results There was a significant correlation between SpT-LS and SN-LS at 30 Hz ( R2 = 0.99; P < .0001), 60 Hz ( R2 = 0.98; P < .0001), and 90 Hz ( R2 = 0.99; P < .0001). There was also a significant correlation between ROT derived from SpT and ROT derived from CT: R2 = 0.97 ( P < .0001) at 30 Hz, R2 = 0.95 ( P < .0001) at 60 Hz, and R2 = 0.98 ( P < .0001) at 90 Hz. Conclusions Point-to-point SpT permits the determination of LS between 2 distinct tissue regions as well as ROT measurement of specific tissue regions without the need for border detection. |
---|---|
ISSN: | 0894-7317 1097-6795 |
DOI: | 10.1016/j.echo.2008.06.004 |