Failure of the hinge mechanism of a trispherical total wrist arthroplasty: A case report and review of the literature

We report a patient with rheumatoid arthritis who developed late catastrophic failure of the hinge mechanism of her trispherical total wrist arthroplasty. This was associated with synovitis secondary to wear debris from Titanium, cement, and polyethylene which produced exuberant flexor and extensor...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of hand surgery (American ed.) 1999-01, Vol.24 (1), p.156-160
Hauptverfasser: O'Flynn, Hugh M., Rosen, Alan, Weiland, Andrew J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report a patient with rheumatoid arthritis who developed late catastrophic failure of the hinge mechanism of her trispherical total wrist arthroplasty. This was associated with synovitis secondary to wear debris from Titanium, cement, and polyethylene which produced exuberant flexor and extensor tendon synovitis and median nerve compression. (J Hand Surg 1999;24A:156–160. Copyright © 1999 by the American Society for Surgery of the Hand.)