Direct determination of stability of protease inhibitors in plasma by HPLC with automated column-switching
Automated column-switching HPLC methods were developed and utilized for the direct analyses of three hydroxamic acid based metalloprotease inhibitors in rat plasma. These column-switching methods involved the use of a restricted-access media (RAM) precolumn and a column-switching valve, allowing the...
Gespeichert in:
Veröffentlicht in: | Journal of pharmaceutical and biomedical analysis 1999-03, Vol.19 (3), p.343-349 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Automated column-switching HPLC methods were developed and utilized for the direct analyses of three hydroxamic acid based metalloprotease inhibitors in rat plasma. These column-switching methods involved the use of a restricted-access media (RAM) precolumn and a column-switching valve, allowing the complete automation of sample preparation and HPLC. The plasma samples were directly injected onto a precolumn packed with SPS/ODS stationary phase and then backflushed onto an ODS analytical column using a 6-port column-switching device. The drug stability in rat plasma was determined using both the automated and traditional HPLC methods. The results obtained from the automated column-switching methods were in good agreement with those from traditional methods that involve sequential protein precipitation, liquid extraction, solvent evaporation, and sample reconstitution. In addition to the elimination of labor-intensive and time-consuming sample preparation procedures, the column-switching methods allowed on-line analyte enrichment and accurate determination of drug stability in plasma with detection limits in the range of 10–20 ng ml
−1. This work represents, for the first time, a drug stability study in plasma by automated column-switching HPLC technique with the use of a RAM column. Our column-switching methods can be readily adapted to any existing HPLC system with minimal hardware modification. |
---|---|
ISSN: | 0731-7085 1873-264X |
DOI: | 10.1016/S0731-7085(98)00127-7 |