In vivo dosimetry using diodes as a quality control tool--experience of 2 years and 2000 patients
This paper reports the use of diodes as a quality control tool on more than 95% of patients undergoing megavoltage radiotherapy during the past 2 years. The paper describes a simple method for using the diodes, a number of possible pitfalls and a quality issue raised by the measurements. In general,...
Gespeichert in:
Veröffentlicht in: | British journal of radiology 1999-09, Vol.72 (861), p.891-895, Article 891 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper reports the use of diodes as a quality control tool on more than 95% of patients undergoing megavoltage radiotherapy during the past 2 years. The paper describes a simple method for using the diodes, a number of possible pitfalls and a quality issue raised by the measurements. In general, the results have been reassuring with very few patients falling outside a +/- 5% action level. |
---|---|
ISSN: | 0007-1285 1748-880X |
DOI: | 10.1259/bjr.72.861.10645196 |