Pseudofracture of an ICD Lead

A proximal pseudofracture of a Medtronic 6936 defibrillating lead adjacent to the ICD pulse generator was found on X ray. At this site, the multifilar metallic connector joins the stimulating lead before coiling around it. Conventional X ray of the abdomen does not visualize this connection. During...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1999-11, Vol.22 (11), p.1704-1704
Hauptverfasser: CAMOUS, JEAN PIERRE, RAYBAUD, FLORENCE, JOURDAN, JACQUES
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A proximal pseudofracture of a Medtronic 6936 defibrillating lead adjacent to the ICD pulse generator was found on X ray. At this site, the multifilar metallic connector joins the stimulating lead before coiling around it. Conventional X ray of the abdomen does not visualize this connection. During ICD replacement, lead impedance should be routinely measured.