Rapid Extraction of Clenbuterol from Human and Calf Urine Using Empore™ C8 Extraction Disks
In the present paper, disk extraction was evaluated for the rapid isolation of clenbuterol from human and calf urine, followed by high-performance liquid chromatography analysis with UV detection. A method was developed for the extraction with standard density C8 disks. The disks could be washed wit...
Gespeichert in:
Veröffentlicht in: | Journal of analytical toxicology 1999-11, Vol.23 (7), p.632-365 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In the present paper, disk extraction was evaluated for the rapid isolation of clenbuterol from human and calf urine, followed by high-performance liquid chromatography analysis with UV detection. A method was developed for the extraction with standard density C8 disks. The disks could be washed with 25% methanol in 0.01M sodium hydroxide without significant losses of clenbuterol. The recovery of clenbuterol was about 85%, and the extracts were clean. The detection limit was about 10 ng/mL. The main advantages of these disks were the saving of time and the reduced amounts of organic solvents needed. |
---|---|
ISSN: | 0146-4760 1945-2403 |
DOI: | 10.1093/jat/23.7.632 |