Calibration model transfer for near-infrared spectra based on canonical correlation analysis

In order to solve the calibration transformation problem in near-infrared (NIR) spectroscopy, a method based on canonical correlation analysis (CCA) for calibration model transfer is developed in this work. Two real NIR data sets were tested. A comparative study between the proposed method and piece...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytica chimica acta 2008-08, Vol.623 (1), p.22-29
Hauptverfasser: Fan, Wei, Liang, Yizeng, Yuan, Dalin, Wang, Jiajun
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In order to solve the calibration transformation problem in near-infrared (NIR) spectroscopy, a method based on canonical correlation analysis (CCA) for calibration model transfer is developed in this work. Two real NIR data sets were tested. A comparative study between the proposed method and piecewise direct standardization (PDS) was conducted. It is shown that the transfer results obtained with the proposed method based on CCA were better than those obtained by PDS when the subset had sufficient samples.
ISSN:0003-2670
1873-4324
DOI:10.1016/j.aca.2008.05.072