An MRI-based technique for assessment of lower extremity deformities—reproducibility, accuracy, and clinical application
The hypothesis of this study was that length, torsion, and axis of a leg phantom can be measured accurately and reproducibly by magnetic resonance imaging (MRI) and that this can be applied to patients with leg deformities. Two phantoms and 30 patients (genu varum, n = 15; genu valgum, n = 15) wer...
Gespeichert in:
Veröffentlicht in: | European radiology 2008-07, Vol.18 (7), p.1497-1505 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The hypothesis of this study was that length, torsion, and axis of a leg phantom can be measured accurately and reproducibly by magnetic resonance imaging (MRI) and that this can be applied to patients with leg deformities. Two phantoms and 30 patients (genu varum,
n
= 15; genu valgum,
n
= 15) were investigated using an optimized MRI technique. Reference measurements were performed with a micrometer screw and a goniometer. Patient leg length and axis were compared with long radiographs in bipedal stance. Intra- and interobserver reproducibility and accuracy were calculated using the mean absolute difference (MAD) and the 95% confidence interval. In patients, comparisons were done using a paired Student’s
t
-test. MAD, intraobserver MAD, and interobserver MAD were 0.03, 0.03, 0.04 mm (length); 0.98, 1.2, 0.98° (torsion); and 0.18, 0.23, 0.22° (axis), respectively. In patients, leg length was underestimated by MRI (−2.4 ± 0.7%; 1.9 ± 0.7 cm;
P
0.05), while it was significantly underestimated in valgus knees (−3.6 ± 2.8°;
P
|
---|---|
ISSN: | 0938-7994 1432-1084 |
DOI: | 10.1007/s00330-008-0896-y |