Effects of design changes on cemented tapered femoral stem fixation

The effects of matte finish and modularity on loosening of tapered stems using the same cementing technique were studied prospectively. In 80 patients, 82 cemented Exeter primary stems were implanted at total hip revision by one surgeon using the same surgical and cementing technique throughout the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical orthopaedics and related research 1998-10, Vol.355 (355), p.47-56
Hauptverfasser: MIDDLETON, R. G, HOWIE, D. W, COSTI, K, SHARPE, P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The effects of matte finish and modularity on loosening of tapered stems using the same cementing technique were studied prospectively. In 80 patients, 82 cemented Exeter primary stems were implanted at total hip revision by one surgeon using the same surgical and cementing technique throughout the series. The polished stems behaved differently than the matte surfaced stems behaved. Polished stems subsided in the cement mantle an average of 1 mm at 2 years after implantation, but without subsequent loosening of stems at as long as 12 years after implantation. Matte surfaced stems with metal centralizers had a higher loosening rate, and loss of fixation at the prosthesis to cement interface was identified as an early sign of loosening of these stems. At a mean 6-year followup, there were no revisions nor was there radiographic evidence of loosening of the polished modular stems. It is concluded that matte finish results in increased loosening of tapered stems but the introduction of modularity did not.
ISSN:0009-921X
1528-1132
DOI:10.1097/00003086-199810000-00006