T-wave oversensing and inappropriate shocks: a case report
A 27-year-old male with congenital long QT syndrome, SCN5A mutation experienced recurrent inappropriate exercise-related implantable cardioverter defibrillator (ICD) shocks. This device showed T-wave oversensing with double, which lead to these device discharges. Dynamic T-wave oversensing was repro...
Gespeichert in:
Veröffentlicht in: | Europace (London, England) England), 2008-05, Vol.10 (5), p.552-555 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A 27-year-old male with congenital long QT syndrome, SCN5A mutation experienced recurrent inappropriate exercise-related implantable cardioverter defibrillator (ICD) shocks. This device showed T-wave oversensing with double, which lead to these device discharges. Dynamic T-wave oversensing was reproducibly provoked at exercise treadmill testing and was confirmed as the mechanism leading to double counting. The insertion of a new pacing and sensing lead with increased R-wave amplitude did not solve the problem. Exchanging the existing ICD generator with one capable of automatic sensitivity control (Biotronik, Lexos DR, Biotronik, Berlin, Germany) completely eliminated T-wave oversensing and inappropriate shocks. |
---|---|