Use of the HemoCue® near patient testing device to measure the concentration of haemoglobin in suction fluid at elective Caesarean section

Summary We have investigated the suitability of the HemoCue® photometer to measure the concentration of haemoglobin in suction fluid obtained at elective caesarean section in 30 women. Laboratory analysis was used as a gold standard against which values generated by the HemoCue were compared. We use...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 2008-05, Vol.63 (5), p.531-534
Hauptverfasser: Gupta, A., Wrench, I. J., Feast, M. J., Alderson, J. D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Summary We have investigated the suitability of the HemoCue® photometer to measure the concentration of haemoglobin in suction fluid obtained at elective caesarean section in 30 women. Laboratory analysis was used as a gold standard against which values generated by the HemoCue were compared. We used the method of Bland and Altman to analyse the data. The bias and the limits of agreement were −0.013 and −0.39 to 0.36 mg.dl−1 respectively, indicating a good level of agreement. Mean (SD) total blood loss calculated using these data, combined with the weight of the swabs, was consistently greater than clinical estimation: 768 (496) ml versus 506 (249) ml respectively (p 