Cardiac resynchronisation therapy as a cause of the electrical storm
We report a case of young patient with dilated cardiomyopathy and implanted cardioverter-defibrillator in which resynchronisation therapy (CRT-D) induced an electrical storm. One month after implantation of a cardiac resynchronisation pacemaker (CRT-P) the patient suffered from ventricular tachycard...
Gespeichert in:
Veröffentlicht in: | Kardiologia polska (1957) 2007-11, Vol.65 (11), p.1350-1353 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | pol |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report a case of young patient with dilated cardiomyopathy and implanted cardioverter-defibrillator in which resynchronisation therapy (CRT-D) induced an electrical storm. One month after implantation of a cardiac resynchronisation pacemaker (CRT-P) the patient suffered from ventricular tachycardia with poor haemodynamic status and was treated by implantation of a CRT-D with a Y adaptor. After replacement of the CRT-D due to Y adaptor damage (new device without a Y adaptor) we observed an electrical storm during ventricular pacing (biventricular, right and left ventricular pacing respectively). Changing pacing mode from DDDR to AAIR resolved ventricular tachycardias in that patient. |
---|---|
ISSN: | 0022-9032 |