A Portable Visual Acuity Device Used in Diabetic Retinal Screening

Purpose To correlate visual acuity measured with the Optec 800 to the gold-standard Early Treatment Diabetic Retinopathy Study (ETDRS) chart. Design Prospective, cross-sectional study. Methods Forty patients of the Vanderbilt Eye Institute agreed to undergo visual acuity testing. Five patients were...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of ophthalmology 2007-03, Vol.143 (3), p.514-515.e1
Hauptverfasser: Kehler, Lori Ann F., OD, Kehler, K. Bradley, OD, Merin, Lawrence M., BA, Tsai, James C., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose To correlate visual acuity measured with the Optec 800 to the gold-standard Early Treatment Diabetic Retinopathy Study (ETDRS) chart. Design Prospective, cross-sectional study. Methods Forty patients of the Vanderbilt Eye Institute agreed to undergo visual acuity testing. Five patients were excluded because of enucleation or functional blindness of one or both eyes. Two patients were excluded because of poor cooperation during testing. Patients were randomized to one of the two methods and participated in visual acuity testing in a forced-choice manner. The sequence was repeated with the alternate instrument. A total-correct score was calculated for each data set. Results There was no statistically significant difference in scores between the two methods ( P1 = .153, P2 = .307.) Other factors such as gender, age, and instrument testing order did not affect performance. Age was inversely correlated with total correct score. Conclusions In this small cohort, the Optec 800 (Chicago, Illinois, USA) and the ETDRS chart recorded similar visual acuity measurements.
ISSN:0002-9394
1879-1891
DOI:10.1016/j.ajo.2006.09.053