Mechanical Evaluation of Fracture Fixation Augmented With Tricalcium Phosphate Bone Cement in a Porous Osteoporotic Cancellous Bone Model
OBJECTIVE:The purpose of this study was to examine the effects of resorbable bone cement on screw and plate-screw fracture fixation in a porous osteoporotic bone model. METHODS:Experiment 1Screw pullout strength was assessed for 4 sets of 4.5-mm cortical screws inserted into a synthetic osteoporotic...
Gespeichert in:
Veröffentlicht in: | Journal of orthopaedic trauma 2007-02, Vol.21 (2), p.124-128 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVE:The purpose of this study was to examine the effects of resorbable bone cement on screw and plate-screw fracture fixation in a porous osteoporotic bone model.
METHODS:Experiment 1Screw pullout strength was assessed for 4 sets of 4.5-mm cortical screws inserted into a synthetic osteoporotic cancellous bone model, including screws inserted without cement augmentation (control), screws augmented with tricalcium phosphate (TCP) bone cement (Norian SRS; Synthes USA, Paoli, PA), and screws augmented with polymethylmethacrylate. Experiment 2The effects of cement augmentation on plate-screw fixation strength were examined by performing cantilever bending tests on 4 sets of 8 plate-screw constructions, including nonaugmented and TCP-augmented standard and locked screw-plate constructions in a similar bone model.
RESULTS:Experiment 1Cement augmentation with both TCP and polymethylmethacrylate increased screw pullout strength from a porous osteoporotic cancellous bone model by about 4-fold (P < 0.05), and there was no significant difference between the 2 cements (P > 0.1). Experiment 2Fixation strength was 1.5 times higher for locked plates compared with standard plates when neither was augmented with cement (P = 0.07). Cement augmentation with TCP improved fixation strength by 3.6 times for a standard plate-screw construction (P < 0.05) and 3.3 times for a locked plate-screw construction (P < 0.05). The most stable construction was the TCP-augmented locked plate, in which a 5-fold increase was observed compared with that of standard plates without TCP (P < 0.05).
CONCLUSIONS:This study indicates augmenting screws with TCP cement during osteosynthesis improves fixation strength in an osteoporotic cancellous bone model.
CLINICAL RELEVANCE:In fracture situations in which osteoporotic bone makes screw and screw-plate fixation tenuous, screw augmentation with TCP cement should be considered as adjunct treatment. |
---|---|
ISSN: | 0890-5339 1531-2291 |
DOI: | 10.1097/BOT.0b013e318033093e |