Stability of lumbar spine after artificial disc replacement
To determine the influence of artificial disc replacement on the stability of lumbar. Ten fresh human cadaveric lumbar spine specimens (9 males, 1 female) were tested. The donors had a mean age of 29.3 years old. The segments were fixed on a special testing device mounted in a universal testing mach...
Gespeichert in:
Veröffentlicht in: | Zhong nan da xue xue bao. Journal of Central South University. Yi xue ban 2006-10, Vol.31 (5), p.728-731 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | chi |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To determine the influence of artificial disc replacement on the stability of lumbar.
Ten fresh human cadaveric lumbar spine specimens (9 males, 1 female) were tested. The donors had a mean age of 29.3 years old. The segments were fixed on a special testing device mounted in a universal testing machine. Physiological load with a maximum of 10 Nm was applied in the flexion, extension, right and left side bending. Moments were applied in 6 load steps: 0, 2, 4, 6, 8, and 10 Nm. Linearity migration and angle migration were recorded,and the mean flexibility coefficients were calculated. Four procedures of measurement were performed in every specimen: intact segments, segments after discectomy, segments after the artificial disc replacement, and segments after the artificial disc replacement with cutting posterior longitudinal ligament locally.
The indexes we calculated of the segments after the discectomy were statistically different compared with the intact segments (P |
---|---|
ISSN: | 1672-7347 |