Investigation into the factors affecting accuracy of mass measurements on a time-of-flight mass spectrometer using Design of Experiment
The results of an investigation of the parameters which have the most significant effect on the accuracy of mass measurements on a quadrupole orthogonal acceleration time‐of‐flight mass spectrometer (q‐oaToF) are reported. The influence of eight factors is investigated: ion abundances of reference a...
Gespeichert in:
Veröffentlicht in: | Rapid communications in mass spectrometry 2007-01, Vol.21 (4), p.529-535 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The results of an investigation of the parameters which have the most significant effect on the accuracy of mass measurements on a quadrupole orthogonal acceleration time‐of‐flight mass spectrometer (q‐oaToF) are reported. The influence of eight factors is investigated: ion abundances of reference and analyte compounds, mass difference between analyte and reference compounds, quality of calibration, number of reference acquisitions averaged and TDC (time‐to‐digital converter) settings (resolution, Np multiplier (number of pushes correction factor), minimum number of points, i.e. minimum acquisition width which defines a peak). To extract the maximum information from as few experiments as possible, a Design of Experiment approach was used. The data will be used as a basis for developing guidance on accurate mass measurement on q‐oaToF instruments. Copyright © 2007 John Wiley & Sons, Ltd. |
---|---|
ISSN: | 0951-4198 1097-0231 |
DOI: | 10.1002/rcm.2852 |