Analysis of Micafungin in Serum by Capillary Zone Electrophoresis

A method is developed for measuring the concentration of micafungin, an echinocandin antifungal agent, in serum, using capillary zone electrophoresis. With a 0.1M borate buffer (pH 10.0) as the run buffer, detection is carried out at 200 nm. Pretreatment is performed by adding acetonitrile to serum...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of chromatographic science 2007-01, Vol.45 (1), p.28-32
Hauptverfasser: Kitahashi, Toshihiro, Furuta, Itaru
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method is developed for measuring the concentration of micafungin, an echinocandin antifungal agent, in serum, using capillary zone electrophoresis. With a 0.1M borate buffer (pH 10.0) as the run buffer, detection is carried out at 200 nm. Pretreatment is performed by adding acetonitrile to serum (1:2) for deproteinization, allowing the supernatant fluid to be taken for measurement. The detection limit of the assay is 0.5 mg/L at a signal-to-noise ratio of 3.0. Coefficients of variation for intra- and inter-assay precision are 3.45–4.47% and 2.61–7.15%, respectively, at a nominal concentration of 5.0–25.0 mg/L. Their recovery rates are 92–110%. It is convenient for the monitoring of micafungin therapy in patients contracting clinically important deep mycoses.
ISSN:0021-9665
1945-239X
DOI:10.1093/chromsci/45.1.28