Strategy for safe performance of extrathoracic magnetic resonance imaging at 1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients : A prospective study with 115 examinations

The purpose of the present study was to evaluate a strategy for safe performance of extrathoracic magnetic resonance imaging (MRI) in non-pacemaker-dependent patients with cardiac pacemakers. Inclusion criteria were presence of a cardiac pacemaker and urgent clinical need for an MRI examination. Pac...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Circulation (New York, N.Y.) N.Y.), 2006-09, Vol.114 (12), p.1285-1292
Hauptverfasser: SOMMER, Torsten, NAEHLE, Claas P, LITT, Harold, SCHILD, Hans, YANG, Alexander, ZEIJLEMAKER, Volkert, HACKENBROCH, Matthias, SCHMIEDEL, Alexandra, MEYER, Carsten, STRACH, Katharina, SKOWASCH, Dirk, VAHLHAUS, Christian
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of the present study was to evaluate a strategy for safe performance of extrathoracic magnetic resonance imaging (MRI) in non-pacemaker-dependent patients with cardiac pacemakers. Inclusion criteria were presence of a cardiac pacemaker and urgent clinical need for an MRI examination. Pacemaker-dependent patients and those requiring examinations of the thoracic region were excluded. The study group consisted of 82 pacemaker patients who underwent a total of 115 MRI examinations at 1.5T. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 1.5 W/kg. All pacemakers were reprogrammed before MRI: If heart rate was 60 bpm, sense-only mode was used to avoid MR-induced competitive pacing and potential proarrhythmia. Patients were monitored with ECG and pulse oximetry. All pacemakers were interrogated immediately before and after the MRI examination and after 3 months, including measurement of pacing capture threshold (PCT) and serum troponin I levels. All MR examinations were completed safely. Inhibition of pacemaker output or induction of arrhythmias was not observed. PCT increased significantly from pre- to post-MRI (P=0.017). In 2 of 195 leads, an increase in PCT was only detected at follow-up. In 4 of 114 examinations, troponin increased from a normal baseline value to above normal after MRI, and in 1 case (troponin pre-MRI 0.02 ng/mL, post-MRI 0.16 ng/mL), this increase was associated with a significant increase in PCT. Extrathoracic MRI of non-pacemaker-dependent patients can be performed with an acceptable risk-benefit ratio under controlled conditions and by taking both MR- and pacemaker-related precautions.
ISSN:0009-7322
1524-4539
DOI:10.1161/CIRCULATIONAHA.105.597013