Influence of monitor luminance and room illumination on soft-copy reading evaluation with electronically generated contrast-detail phantom: comparison of cathode-ray tube monitor with liquid crystal display

The influence of monitor brightness and room illumination on soft-copy diagnosis by both cathode-ray tube (CRT) monitor and liquid crystal display (LCD) was evaluated and compared using a contrast-detail phantom. Nine observers (7 radiologists and 2 radiological technicians) interpreted six types of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nagoya journal of medical science 2006-06, Vol.68 (3-4), p.115-120
Hauptverfasser: Muramoto, Hideyuki, Shimamoto, Kazuhiro, Ikeda, Mitsuru, Koyama, Kazuyuki, Fukushima, Hiromichi, Ishigaki, Takeo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The influence of monitor brightness and room illumination on soft-copy diagnosis by both cathode-ray tube (CRT) monitor and liquid crystal display (LCD) was evaluated and compared using a contrast-detail phantom. Nine observers (7 radiologists and 2 radiological technicians) interpreted six types of electronically generated contrast-detail phantom images using a 21-inch CRT (2,048x2,560) and a 21-inch LCD (2,048x2,560) under 6 kinds of viewing conditions, i.e. monitor brightness of 330 cd/m2 or 450 cd/m2, and room illumination of 20, 100 or 420 lux at the center of the display. Observers were requested to determine the visible borderline of the objects. Between 330 cd/m2 and 450 cd/m2, no significant difference in the visible area was found under any of the three lighting conditions. However, in two low-contrast phantom images, the visible area on the LCD was significantly larger than that on the CRT, independent of both monitor brightness and room illumination. (p