A Comparison of Device Measures of Dorsal First Ray Mobility

Background: Devices built by Glasoe and Klaue have been used in several studies to measure first ray mobility. Both devices measure sagittal motion of the first ray in a dorsal direction. The primary difference in the devices is the method of the load imposed. This study investigates whether first r...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Foot & ankle international 2005-11, Vol.26 (11), p.957-961
Hauptverfasser: Glasoe, Ward M., Grebing, Brett R., Beck, Susan, Coughlin, Michael J., Saltzman, Charles L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background: Devices built by Glasoe and Klaue have been used in several studies to measure first ray mobility. Both devices measure sagittal motion of the first ray in a dorsal direction. The primary difference in the devices is the method of the load imposed. This study investigates whether first ray mobility measured with the Glasoe device is similar to the amount of mobility measured with the Klaue device. Methods: Using the devices described by Glasoe and Klaue, dorsal first ray mobility was measured in 39 patients who had foot and ankle problems. Paired t-tests were computed to assess for differences between device measures of dorsal mobility. Intraclass correlation coefficient (ICC) and absolute difference values were computed to further assess the agreement in measures. Results: Dorsal mobility measured with the Glasoe device averaged 4.9 mm (1.8 to 9.3 mm). Dorsal mobility measured with the Klaue device averaged 5.2 mm (2.5 to 8.5 mm). Paired t-tests (p = 0.12) revealed no significant difference in measures. An ICC of 0.70 and a mean absolute difference of 0.9 mm (SD 0.8) were found between the two clinical measures further suggesting agreement. Conclusion: Results indicated that the two devices possess similar diagnostic accuracy in the measurement of dorsal first ray mobility.
ISSN:1071-1007
1944-7876
DOI:10.1177/107110070502601111