A comparison of cyclic fatigue between used and new Mtwo Ni-Ti rotary instruments
Aim To evaluate the cyclic fatigue of Mtwo Ni–Ti rotary instruments after controlled clinical use in molar teeth. Methodology Twenty Mtwo instruments of each size were selected and divided into two groups: group A consisted of 10 new instruments (control group); group B consisted of 10 used instru...
Gespeichert in:
Veröffentlicht in: | International endodontic journal 2006-09, Vol.39 (9), p.716-723 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aim To evaluate the cyclic fatigue of Mtwo Ni–Ti rotary instruments after controlled clinical use in molar teeth.
Methodology Twenty Mtwo instruments of each size were selected and divided into two groups: group A consisted of 10 new instruments (control group); group B consisted of 10 used instruments. Each instrument in group B was used to clean and shape 10 root canals of molar teeth in patients. Cyclic fatigue testing of instruments was performed in tapered artificial canals with a 5‐mm radius of curvature and a 60° angle of curvature. In all 140, instruments were rotated until fracture and the number of cycles to failure was recorded. Data were analysed using one‐way anova, Tukey's HSD test and independent sample t‐test to determine any statistical difference; the significance was determined at the 95% confidence level.
Results A reduction of cycles to failure between new (group A) and used (group B) instruments was apparent. A statistically significant difference (P |
---|---|
ISSN: | 0143-2885 1365-2591 |
DOI: | 10.1111/j.1365-2591.2006.01142.x |